New TopShim run:

version = topshim 2.2.0
date = 2024-06-18
time = 18:21:46.825
command = topshim 

FINE SHIMMING

Parameters:

probe = Z163739_0737
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 0.7 Hz, Target= 0.6 Hz, After= 1.8 Hz
Iteration 2: FWHM Before= 1.8 Hz, Target= 0.4 Hz, After= 0.6 Hz
Iteration 3: FWHM Before= 0.6 Hz, Target= 0.5 Hz, After= 0.7 Hz
Iteration 4: FWHM Before= 0.7 Hz, Target= 0.7 Hz, After= 0.7 Hz
Iteration 5: FWHM Before= 0.7 Hz, Target= 0.5 Hz, After= 0.5 Hz
Iteration 6: FWHM Before= 0.5 Hz, Target= 0.3 Hz, After= 0.7 Hz
Iteration 7: FWHM Before= 0.7 Hz, Target= 0.4 Hz, After= 0.7 Hz
Iteration 8: FWHM Before= 0.7 Hz, Target= 0.5 Hz, After= 0.9 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D shimming, non-spinning) = 0.675

shim changes: 
Z	-1026
Z2	-266
Z3	-355
Z4	-560
Z5	-553
Z6	-4239
Z7	-7854
Z8	-1485

duration = 3 min 51 sec
completed successfully
finished Tue Jun 18 18:25:35 2024


========================================================================